GE PT878 Manual do Utilizador Página 186

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 297
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 185
Chapter 8. Servicing the PT878
170 Transport® Model PT878 Portable Liquid Flowmeter User’s Manual
8.4.5 Calculating Velocity (Pipe Material Soundspeed) (cont.)
7. The “Calculated” box shows the thickness value measured. The PT878
asks for confirmation of the calculated and current values. Press
[ENTER] to commit the calculated value, or to recalculate the value.
Note: If you commit the calculated value, the settings in the Material
window also change from the previous material to “Other” with the
new soundspeed.
To leave the Velocity window:
Press the [W] or [X] arrow key to move to another tab.
Press [F2] (Cancel) to return to Operate Mode without confirming the
soundspeed value.
Press [F3] (OK) to confirm the new value and return to Operate
Mode.
8.5 Programming the Thickness Gauge
While the T-Gauge Display option allows you to program the appropriate
material and soundspeed, the Thickness Gauge Setup option enables you to
view or change five parameters:
Low Signal Threshold
Transducer Delay
Signal Inversion
Noise Threshold
Detection Threshold
Note: The thickness gauge programming settings are entered at the factory.
You should not change them unless instructed by the factory.
The Programming window appears similar to Figure 115 on page 171.
Vista de página 185
1 2 ... 181 182 183 184 185 186 187 188 189 190 191 ... 296 297

Comentários a estes Manuais

Sem comentários