GE PT878 Manual do Utilizador Página 64

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 297
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 63
Chapter 3. Programming Site Data
48 Transport® Model PT878 Portable Liquid Flowmeter User’s Manual
3.7.2 Entering Custom Cp Data in the Energy Option (cont.)
2. The next prompt asks from which source — a fixed value or table —
the PT878 will use for fluid density values. Use the [
W] and [X] keys
to scroll to the appropriate radio button, and press
[ENTER].
3. The menu now varies, depending on your selection in step 2.
If you selected “Fixed,” the PT878 asks for the fixed fluid density.
Use the numeric keys to enter the desired value, and press
[ENTER].
If you selected “Table,” the meter highlights the “Edit Table” button.
Figure 25: Fluid Density Table
a. Press [ENTER] to open the Edit Density table, shown in Figure 25
above.
b. Use the numeric keys to enter the desired temperature in degrees
Kelvin, and press
[ENTER].
c. Press the [X] key to move to the kg/m^3 column (density in
Kilogram/cubic meters). Use the numeric keys to enter the desired
value, and press
[ENTER].
Edit Density
Cancel OK
°Kelvin kg/m^3
1
2
3
4
5
Vista de página 63
1 2 ... 59 60 61 62 63 64 65 66 67 68 69 ... 296 297

Comentários a estes Manuais

Sem comentários