Coupling
Coupling
Coupling is used when adjusting the coupling between the electronic head and the torque
tube. To adjust:
1. Click Coupling Method and the coupling appears and is updated.
2. Adjust the physical coupling until this value is between +/- 5%.
3. Click Coupling Method a second time to turn off the coupling measurement.
Menu - Calibrate
When you right-click on any of the Calibrate tabs, a context menu appears. The items that
appear depend on the HDLT mode:
Report Setup - Allows you to select a report file and a report printing program.
Report - Generates a report of the HDLT parameters.
Detach Trend - Removes the Trend display from the tabbed dialog and creates a
separate trend display.
Help - Displays the help file at the Monitor tab instructions.